P/N: 036294900
Standard version with coated vertical glassware: Version for Hei-VAP Benchtop, including G9B glassware
Using the control panel on the control box of the Hei-VOLUME Distimatic module, you can enter the parameters for the sensor or time-controlled filling of the rotary flask, as well as program the emptying time of the coated collector vessel (approx. 400 ml) using the chemical membrane pump. The control box can be mounted on a nearby wall using the supplied bracket with hose guide.
In combination with the corresponding evaporator system, the Hei-VOLUME Distimatic automatic module enables efficient, unattended operation for every rage of application.
For the greatest possible safety and a wide range of applications, all parts that come into contact with the media, including the supplied tubing, are made of chemical-resistant materials.
P/N: 036294901
For high performance distillation, suitable for most solvents: Version for Hei-VAP Benchtop, including G9B XL glassware
Using the control panel on the control box of the Hei-VOLUME Distimatic module, you can enter the parameters for the sensor or time-controlled filling of the rotary flask, as well as program the emptying time of the coated collector vessel (approx. 400 ml) using the chemical membrane pump. The control box can be mounted on a nearby wall using the supplied bracket with hose guide.
In combination with the corresponding evaporator system, the Hei-VOLUME Distimatic automatic module enables efficient, unattended operation for every rage of application.
For the greatest possible safety and a wide range of applications, all parts that come into contact with the media, including the supplied tubing, are made of chemical-resistant materials.
Using the control panel on the control box of the Hei-VOLUME Distimatic module, you can enter the parameters for the sensor or time-controlled filling of the rotary flask, as well as program the emptying time of the coated collector vessel (approx. 400 ml) using the chemical membrane pump. The control box can be mounted on a nearby wall using the supplied bracket with hose guide.
In combination with the corresponding evaporator system, the Hei-VOLUME Distimatic automatic module enables efficient, unattended operation for every rage of application. The following versions are available for you to choose from:
In the standard version with coated vertical glassware: Version for Hei-VAP Benchtop, including G9B glassware
For high performance distillation, suitable for most solvents: Version for Hei-VAP Benchtop, including G9B XL glassware
For the greatest possible safety and a wide range of applications, all parts that come into contact with the media, including the supplied tubing, are made of chemical-resistant materials.
Compatibility | Hei-VAP and Laborota 4000 series |
---|---|
Automatic refill of sample | yes |
Automatic discharge of condensate | yes |
Automatic drainage of residue | no |
Power input | 3,500 W |
Emergency-Stop button | yes |
Protection class IEC 60529 | IP 65 (control panel and residue pump) IP 54 (Housing) |
Hei-VOLUME-Certificate-for-continuous-operation.pdf | Download |
Hei-VOLUME Distimatic - The clever way of continuous evaporation | Download |
Operation Manual-Distimatic-EN.pdf | Download |
Installation-Instruction_Valve-matrix-and-Collector.pdf | Download |
Hei-VOLUME Distimatic at a glance | Download |
AWARD-WINNING EVAPORATION SYSTEMS – Rotary evaporators | Download |
Operation Manual-Distimatic-EN.pdf | Download |
Due to the automated feed of only a small amount of medium in the rotary flask, the conditions in the overall system are constant in terms of …
The best way to clean a system with the Hei-VOLUME Distimatic module is …
For the Hei-VOLUME Distimatic Module, the value is 500 mPAs.
There are two modes available for programming the automatic module Hei-VOLUME Distimatic: …
The automatic module Hei-VOLUME Distimatic Benchtop is compatible with …
No, media with solids content cannot be processed with the Hei-VOLUME Distimatic automatic module.
If you have a Hei-VOLUME Distimatic automatic module, you must …
Yes. Thanks to sensor monitoring and emergency shutdown, …